WaveMaster 813Zi

13 GHz, 40 GS/s, 4 Ch, 10 Mpts/Ch WaveMaster with 15.3" WXGA Color Display. 50 ohm and 1 Mohm Input


WaveMaster 804Zi   (4 GHz Bandwidth, 4 Input Channels, 40 GS/s on 4 Ch (80 GS/s on 2 Ch) Max Sample Rate)
Quick Specs
Analog (ProLink Input) Bandwidth @ 50 Ω (-3 dB)4 GHz
(≥10 mV/div)
Input Channels4 (Any combination of ProLink and ProBus inputs)
Single-Shot Sample Rate/Ch40 GS/s on 4 Ch (80 GS/s on 2 Ch using optional WM8Zi-2X80GS External Interleaving Device)
Standard Memory (4 Ch / 2 Ch) (Number of Segments)10 M / 10 M
(optional WM8Zi-2X80GS External Interleaving Device permits 80 GS/s on 2ch with twice the memory)
(5000)
WaveMaster 806Zi   (6 GHz Bandwidth, 4 Input Channels, 40 GS/s on 4 Ch (80 GS/s on 2 Ch) Max Sample Rate)
Quick Specs
Analog (ProLink Input) Bandwidth @ 50 Ω (-3 dB)6 GHz
(≥10 mV/div)
Input Channels4 (Any combination of ProLink and ProBus inputs)
Single-Shot Sample Rate/Ch40 GS/s on 4 Ch (80 GS/s on 2 Ch using optional WM8Zi-2X80GS External Interleaving Device)
Standard Memory (4 Ch / 2 Ch) (Number of Segments)10 M / 10 M
(optional WM8Zi-2X80GS External Interleaving Device permits 80 GS/s on 2ch with twice the memory)
(5000)
WaveMaster 808Zi   (8 GHz Bandwidth, 4 Input Channels, 40 GS/s on 4 Ch (80 GS/s on 2 Ch) Max Sample Rate)
Quick Specs
Analog (ProLink Input) Bandwidth @ 50 Ω (-3 dB)8 GHz
(≥10 mV/div)
Input Channels4 (Any combination of ProLink and ProBus inputs)
Single-Shot Sample Rate/Ch40 GS/s on 4 Ch (80 GS/s on 2 Ch using optional WM8Zi-2X80GS External Interleaving Device)
Standard Memory (4 Ch / 2 Ch) (Number of Segments)10 M / 10 M
(optional WM8Zi-2X80GS External Interleaving Device permits 80 GS/s on 2ch with twice the memory)
(5000)
WaveMaster 813Zi   (13 GHz Bandwidth, 4 Input Channels, 40 GS/s on 4 Ch (80 GS/s on 2 Ch) Max Sample Rate)
Quick Specs
Analog (ProLink Input) Bandwidth @ 50 Ω (-3 dB)13 GHz
(≥10 mV/div)
Input Channels4 (Any combination of ProLink and ProBus inputs)
Single-Shot Sample Rate/Ch40 GS/s on 4 Ch (80 GS/s on 2 Ch using optional WM8Zi-2X80GS External Interleaving Device)
Standard Memory (4 Ch / 2 Ch) (Number of Segments)10 M / 10 M
(optional WM8Zi-2X80GS External Interleaving Device permits 80 GS/s on 2ch with twice the memory)
(5000)
WaveMaster 816Zi   (16 GHz Bandwidth, 4 Input Channels, 40 GS/s on 4 Ch (80 GS/s on 2 Ch) Max Sample Rate)
Quick Specs
Analog (ProLink Input) Bandwidth @ 50 Ω (-3 dB)16 GHz
(≥10 mV/div)
Input Channels4 (Any combination of ProLink and ProBus inputs)
Single-Shot Sample Rate/Ch40 GS/s on 4 Ch (80 GS/s on 2 Ch using optional WM8Zi-2X80GS External Interleaving Device)
Standard Memory (4 Ch / 2 Ch) (Number of Segments)10 M / 10 M
(optional WM8Zi-2X80GS External Interleaving Device permits 80 GS/s on 2ch with twice the memory)
(5000)
WaveMaster 820Zi   (16 GHz Bandwidth, 4 Input Channels, 40 GS/s on 4 Ch (80 GS/s on 2 Ch) Max Sample Rate)
Quick Specs
Analog (ProLink Input) Bandwidth @ 50 Ω (-3 dB)16 GHz
(≥10 mV/div)
Input Channels4 (Any combination of 16 GHz ProLink inputs or 3.5 GHz ProBus inputs),
3 (1 @ full BW, 2 with ProLink or ProBus inputs), or
2 (@ full BW)
Single-Shot Sample Rate/Ch40 GS/s on 4 Ch (80 GS/s on 2 Ch when operated in ≥20 GHz Mode)
Standard Memory (4 Ch / 2 Ch) (Number of Segments)10 M / 10 M
(20 Mpts in 2 Ch mode when operated in ≥20 GHz Mode)
(5000)
WaveMaster 825Zi   (16 GHz Bandwidth, 4 Input Channels, 40 GS/s on 4 Ch (80 GS/s on 2 Ch) Max Sample Rate)
Quick Specs
Analog (ProLink Input) Bandwidth @ 50 Ω (-3 dB)16 GHz
(≥10 mV/div)
Input Channels4 (Any combination of 16 GHz ProLink inputs or 3.5 GHz ProBus inputs),
3 (1 @ full BW, 2 with ProLink or ProBus inputs), or
2 (@ full BW)
Single-Shot Sample Rate/Ch40 GS/s on 4 Ch (80 GS/s on 2 Ch when operated in ≥20 GHz Mode)
Standard Memory (4 Ch / 2 Ch) (Number of Segments)10 M / 10 M
(20 Mpts in 2 Ch mode when operated in ≥20 GHz Mode)
(5000)
WaveMaster 830Zi   (16 GHz Bandwidth, 4 Input Channels, 40 GS/s on 4 Ch (80 GS/s on 2 Ch) Max Sample Rate)
Quick Specs
Analog (ProLink Input) Bandwidth @ 50 Ω (-3 dB)16 GHz
(≥10 mV/div)
Input Channels4 (Any combination of 16 GHz ProLink inputs or 3.5 GHz ProBus inputs),
3 (1 @ full BW, 2 with ProLink or ProBus inputs), or
2 (@ full BW)
Single-Shot Sample Rate/Ch40 GS/s on 4 Ch (80 GS/s on 2 Ch when operated in ≥20 GHz Mode)
Standard Memory (4 Ch / 2 Ch) (Number of Segments)10 M / 10 M
(20 Mpts in 2 Ch mode when operated in ≥20 GHz Mode)
(5000)
Check up to four models. Compare
Innovative, Superior, Upgradeable

LeCroy has broken bandwidth barriers by combining proven SiGe high-speed components with 5th generation Digital Bandwidth Interleave (DBI) technology to achieve unprecedented real-time oscilloscope performance: 30 GHz, 80 GS/s, and 512 Mpts/Ch Analysis Memory with superior signal fidelity.

5th Generation Digital Bandwidth Interleave (DBI)

As memory and sample rate can be interleaved, so can bandwidth. Using high- performance microwave and RF technologies, and high-speed processors and digital signal processing (DSP), LeCroy intelligently operates high-speed analog components comfortably within their rated bandwidth range while providing twice the four channel bandwidth on two channels. This approach offers better signal fidelity compared to “stretching” of components beyond their rated bandwidth.
See the Additional Resources for more information

4 channels @ 30 GHz with Zi-8CH-SYNCH more info

LeCroy has extended the power of it’s 30 GHz oscilloscope by providing a simple and fast method to combine two WaveMaster 830Zi oscilloscopes and provide 4 channels at 30 GHz. This solution is ideal for measurement and analysis of 28 to 56 Gb/s IQ modulated signals where ultra-high real-time bandwidth and four channels is required, or for capture and detailed analysis of other leading edge technologies [See Additional Resources for more information]. The solution is compatible with all WaveMaster 8 Zi models, so it may be used to create 20 to 30 GHz four channel oscilloscope captures, or 4 to 16 GHz eight channel oscilloscope captures.

Superior Serial Data Analysis

SDA II Serial Data Analysis Software provides the highest confidence for serial data testing. Eye diagram analysis provides 100x faster capture of unit intervals (UI) in the eye diagram, and is enhanced with superior eye diagram analysis tools, such as lines of constant bit error rate (IsoBER ) and mask violation locators. Superior jitter decomposition methodologies and analysis tools provides more ability to understand problem behaviors in serial data systems. X-Stream II architecture leverages the superior tools through 10–100x faster processing and complete analysis processing of full record lengths, providing more insight faster.

Best Investment Protection

From the perspective of bandwidth, sample rate, processing speed, responsiveness, display size, and range of capability, the WaveMaster 8 Zi platform is clearly superior and will remain so for many years to come. With the widest bandwidth upgrade range, an engineer who is working on 2nd generation technologies today can confidently know that WaveMaster 8 Zi will support 3rd or 4th generation technology needs several years from now.

Most Advanced Oscilloscope Platform

X-Stream II architecture provides 10–100x faster processing of long records, true 512 Mpts analysis memory, and instantaneous instrument responsiveness. When off-line data analysis is preferred, the LeCroy Serial Interface Bus (LSIB) provides up to 325 MB/s data transfer speeds. A 15.3" WXGA touchscreen display provides 50% more viewing area compared to other oscilloscopes in this class. Both 50 Ω and 1 MΩ inputs are standard on all oscilloscopes, as is support for all LeCroy probes(including passive probes) without the need for easy to lose external adapters or expensive power supplies.

Eye Doctor™ II Advanced Signal Integrity Tools Add Measurement Precision
  • Test Fixture and Cable De-embedding
  • Emphasis Emulation
  • Serial Data Channel Emulation
  • Receiver Equalization

Add precision to signal integrity measurements by allowing the subtraction of fixture effects, and the emulation of emphasis, serial data channels, and receiver DFE, FFE, and CTLE equalization effects while at the same time maintaining fast oscilloscope processing speed on unlimited record lengths. Re-capture design margin, better understand actual circuit performance, and perform compliance testing on emerging high-speed standards, such as PCIe Gen3, USB 3.0, SAS/SATA 6 Gb/s, that require emulation to ensure high field reliability and interoperability.

X-Stream II Architecture

 

Optimized for Fast Throughput

X-Stream II architecture enables high throughput of data—even when the oscilloscope is performing multiple 100 Mpts (or larger) waveforms. X-Stream II uses variable waveform segment lengths to enable all processing intensive calculations to take place in fast CPU cache memory, thus improving calculation speed and efficiency. The result—10–100x faster processing compared to other oscilloscopes.

Optimized for Long Memory

X-Stream II essentially has no analysis memory length restrictions, regardless of analysis type, since the variable waveform segment length can always be limited to a size that can fit in CPU cache memory. Other oscilloscopes with conventional architectures cannot make this claim, and often have limitations on analysis memory of 5–20% the length of their acquisition memory under the best conditions.

Optimized for Responsiveness

By dynamically allocating buffers to maximize memory availability, the WaveMaster 8 Zi Series embodies the fastest front panel responsiveness. A built-in processing abort makes front panel control changes instant by stopping the current process and allowing new waveforms to be positioned or zoomed—all without a lengthy recalculation. Meanwhile, waveform previewing shows interim calculation results.

Freedom from Probing Limitations

WaveMaster 8 Zi excels in the way it offers general purpose utility never before seen in oscilloscopes from 4 to 30 GHz. All WaveMaster 8 Zi oscilloscopes contain selectable 50 Ω and 1 MΩ input capability and can be used with any LeCroy probe—passive or active—without requiring external adapters or power supplies.Use the two inputs per channel as a built-in switch matrix.

Complete System Debug

Understanding the relationships between different signals is vital to fast debug. Only WaveMaster 8 Zi combines the best of general purpose oscilloscopes (low-speed serial triggers and decoders, mixed signal capability, high impedance probing)with specialized 50 Ω inputs. This provides ability to 100% correlate high bandwidth signal activities or problems to low-speed events, such as low-speed serial data control words, power supply noise, or parallel data transmissions. Supported serial standards are 8b10b, PCIe, PCIe 2.0, SAS, SATA, XAUI, I2C, SPI CAN, LIN, UART-RS-232, and FlexRay.

More Triggers Isolate More Problems More Effectively

A full range of serial triggers (I2C, SPI, UART, RS-232, CAN, LIN, FlexRay) are available, providing capability to isolate events related to serial peripheral bus traffic. Most serial triggers contain powerful conditional DATA triggering to allow inclusion or exclusion of entire ranges of data values, which expedites understanding of how a range of serial bus traffic DATA values interact with other signals.

TriggerScan™

TriggerScan uses high-speed hardware triggering capability with persistence displays to capture only the signals of interest and provide answers up to 100x faster than other methods.

WaveScan™ Advanced Search and Analysis Finds Problems that Triggers Won’t Find

The best trigger won’t find all unusual events—a more powerful capability is sometimes needed. WaveScan provides the ability to locate unusual events in a single capture (i.e., capture and search) or “scan” for an event in many acquisitions over a long period of time. Select from more than 20 search modes (frequency,rise time, runt, duty cycle, etc.), apply a search condition and begin scanning. When an event is found, WaveScan highlights the error on screen and displays a table listing the errors.

Data Transfer Speeds 25–100x Faster

LeCroy’s Serial Interface Bus (LSIB) option enables direct connection to the PCI Express® x4 high-speed data bus in the oscilloscope to enable data transfer rates up to 325 MB/s. All that is required is installation of an optional LSIB card in the oscilloscope and the corresponding host board (card) for desktop (laptop) PC in the remote computer. Data transfer is easily enabled through a supplied application program interface (API).

Application Specific Solutions

In addition to the general purpose WaveShape Analysis tools, application specific solutions are available for Serial Data Compliance, Embedded Design, Digital Design, and Automotive. These packages extend the LeCroy standard measurement and analysis capabilities and expand your oscilloscope’s utility as your needs change.

  • Digital Filter Software Package (WM8Zi-DFP2)
  • Spectrum Analyzer Analysis Package (WM8Zi-SPECTRUM)
  • Mixed Signal Oscilloscope Option (MS-250/MS-500)
  • Jitter and Timing Analysis Package (WM8Zi-JTA2)
  • Eye Doctor II – Advanced Signal Integrity Tools (WM8Zi-EYEDRII)

Vertical System

Analog (ProLink Input) Bandwidth @ 50 Ω (-3 dB) 13 GHz
(≥10 mV/div)
Analog (ProBus Input) Bandwidth @ 50 Ω (-3 dB) 3.5 GHz
(≥10 mV/div)
Analog (ProBus Input) Bandwidth @ 1 MΩ (-3 dB) 500 MHz (typical, ≥2 mV/div)
Rise Time (typical, 10-90%, 50 Ω) 32.5 ps
(test limit, flatness mode)
Rise Time (typical, 20-80%, 50 Ω) 24.5 ps
(flatness mode)
Input Channels 4 (Any combination of ProLink and ProBus inputs)
Bandwidth Limiters 20 MHz, 200 MHz, 1 GHz, 4 GHz, 6 GHz, 8 GHz
Input Impedance ProLink Inputs: 50 Ω+/-2% for ≤100 mV/div, 50 Ω+/-3% for >100 mV/div
ProBus Inputs: 50 Ω+/-2% or 1 MΩ||16pF, 10 MΩ || 11 pF with supplied Probe
Input Coupling ProLink Inputs - 50 Ω: DC, GND ProBus Inputs - 1 MΩ: AC, DC, GND; 50 Ω: DC, GND
Maximum Input Voltage 50 Ω (ProLink): ±2 Vmax@≤100mV/div, 5.5Vrms@>100mV/div
50 Ω (ProBus): ±5 Vmax, 3.5 Vrms
1 MΩ (ProBus): 250 V max. (peak AC: < 10 kHz + DC)
Channel-Channel Isolation DC to 10 GHz: 50 dB (>315:1)
10 to 15 GHz: 46 dB (>200:1)
15 to 16 GHz: 40 dB (>100:1)
(For any two ProLink input channels, same or different v/div settings, typical)
Vertical Resolution 8 bits; up to 11 bits with enhanced resolution (ERES)
Sensitivity 50 Ω (ProLink): 2 mV–1 V/div, fully variable (2-9.9 mV/div via zoom)
50 Ω (ProBus): 2 mV–1 V/div, fully variable; 1 MΩ (ProBus): 2 mV–10 V/div, fully variable
DC Gain Accuracy (±1.5% of full scale
Offset Range 50 Ω (ProLink):
(±500 mV @ 2–100 mV/div
(±4 V @ >100 mV/div -1 V/div
50 Ω (ProBus):
(±750 mV @ 2–100 mV/div
(±4 V @ >100 mV/div -1 V/div
1 MΩ:
(±1V @ 2–140 mV/div
(±10V @ 142mV-1.40V/div
(±100V @ 1.42V-10V/div
Offset Accuracy (±(1.5% of full scale + 1.5% of offset value + 2 mV)
 

Horizontal System

Timebases Internal timebase common to 4 input channels; an external clock may be applied at the auxiliary input
Time/Division Range 20 ps/div–128 s/div (Real-Time Mode: 20 ps/div - 64 s/div; RIS mode: 20 ps/div - 10 ns/div; Roll mode: 100 ms/div up to 128 s/div), depending on memory length
Clock Accuracy < 1 ppm + (aging of 0.5ppm/yr from last calibration)
Time Interval Accuracy < 0.06 / SR + (clock accuracy* Reading) (rms)
Jitter Noise Floor For Acq. Length ≤10μs: 375 fsrms (TIE, typical)
For Acq. Length >10μs: 425 fsrms (TIE, typical)
Trigger Jitter 2 ps rms (typical)
<0.1 ps rms (typical, software assisted)
Channel-Channel Deskew Range (±9 x time/div. setting or 25 ns max. (whichever is larger), each channel
External Timebase Reference (Input) 10 MHz; 50 Ω impedance, applied at the rear input
External Timebase Reference (Output) 10 MHz; 50 Ω impedance, output at the rear
External Sample Clock Not Available
 

Acquisition System

Single-Shot Sample Rate/Ch 40 GS/s on 4 Ch (80 GS/s on 2 Ch using optional WM8Zi-2X80GS External Interleaving Device)
Random Interleaved Sampling (RIS) 200 GS/s for repetitive signals (20 ps/div to 10 ns/div)
Maximum Trigger Rate 1,000,000 waveforms/second (in Sequence Mode, up to 4 channels)
Intersegment Time 1 μs
Standard Memory (4 Ch / 2 Ch) (Number of Segments) 10 M / 10 M
(optional WM8Zi-2X80GS External Interleaving Device permits 80 GS/s on 2ch with twice the memory)
(5000)
S - Memory Option 32 M / 32 M
(optional WM8Zi-2X80GS External Interleaving Device permits 80 GS/s on 2ch with twice the memory)
(15,000)
M - Memory Option 64 M / 64 M
(optional WM8Zi-2X80GS External Interleaving Device permits 80 GS/s on 2ch with twice the memory)
(15,000)
L - Memory Option 128 M / 128 M
(optional WM8Zi-2X80GS External Interleaving Device permits 80 GS/s on 2ch with twice the memory)
(15,000)
VL – Memory Option 256 M / 256 M
(optional WM8Zi-2X80GS External Interleaving Device permits 80 GS/s on 2ch with twice the memory)
(15,000)
 

Acquisition Modes

Averaging Summed averaging to 1 million sweeps; continuous averaging to 1 million sweeps
Enhanced Resolution (ERES) From 8.5 to 11 bits vertical resolution
Envelope (Extrema) Envelope, floor, or roof for up to 1 million sweeps
Interpolation Linear or Sin x/x
 

Triggering System

Modes Normal, Auto, Single, and Stop
Sources Any input channel, Aux, Aux/10, Line, or Fast Edge. Slope and level unique to each source (except line trigger)
Coupling DC, AC, HFRej, LFRej
Pre-trigger Delay 0–100% of memory size (adjustable in 1% increments of 100 ns)
Post-trigger Delay 0–10,000 divisions in real time mode, limited at slower time/div settings or in roll mode
Hold-off by Time From 2 ns up to 20 s or from 1 to 99,999,999 events
Internal Trigger Range (±4.1 div from center
Trigger Sensitivity with Edge Trigger (Ch 1–4) ProBus Inputs 2 div @ < 3.5 GHz
1.5 div @ < 1.75 GHz
1.0 div @ < 200 MHz
(for DC coupling, ≥ 10 mV/div, 50 Ω )
Trigger Sensitivity with Edge Trigger (Ch 1–4) ProLink Inputs 3 div @ < 13 GHz
1.5 div @ < 3 GHz
1.0 div @ < 200 MHz
(for DC, AC, LFRej coupling, ≥ 10 mV/div, 50 Ω )
External Trigger Sensitivity, (Edge Trigger) 2 div @ < 1 GHz
1.5 div @ < 500 MHz
1.0 div @ < 200 MHz
(for DC coupling)
External Trigger Input Range Aux (±0.4 V); Aux/10 (±4 V)
Max. Trigger Frequency, SMART Trigger 2.0 GHz @ ≥ 10 mV/div (minimum triggerable width 200 ps)
 

Basic Triggers

Edge Triggers when signal meets slope (positive, negative, or either) and level condition.
Window Triggers when signal exits a window defined by adjustable thresholds
TV-Composite Video Triggers NTSC or PAL with selectable line and field; HDTV (720p, 1080i, 1080p) with selectable frame rate (50 or 60 Hz) and Line; or CUSTOM with selectable Fields (1-8), Lines (up to 2000), Frame Rates (25, 30, 50, or 60 Hz), Interlacing (1:1, 2:1, 4:1, 8:1), or Synch Pulse Slope (Positive or Negative).
 

SMART Triggers

State or Edge Qualified Triggers on any input source only if a defined state or edge occurred on another input source. Holdoff between sources is selectable by time or events.
Qualified First In Sequence acquisition mode, triggers repeatably on event B only if a defined pattern, state, or edge (event A) is satisfied in the first segment of the acquisition. Holdoff between sources is selectable by time or events.
Dropout Triggers if signal drops out for longer than selected time between 1 ns and 20 s.
Pattern Logic combination (AND, NAND, OR, NOR) of 5 inputs (4 channels and external trigger input). Each source can be high, low, or don’t care. The High and Low level can be selected independently. Triggers at start or end of the pattern.
 

SMART Triggers with Exclusion Technology

Glitch Triggers on positive or negative glitches with widths selectable as low as 200ps to 20 s, or on intermittent faults.
Width (Signal or Pattern) Triggers on positive, negative, or both widths with widths selectable as low as 200ps to 20 s, or on intermittent faults.
Interval (Signal or Pattern) Triggers on intervals selectable between 1 ns and 20 s.
Timeout (State/Edge Qualified) Triggers on any source if a given state (or transition edge) has occurred on another source. Delay between sources is 1 ns to 20 s, or 1 to 99,999,999 events.
Runt Trigger on positive or negative runts defined by two voltage limits and two time limits. Select between 1 ns and 20 ns.
Slew Rate Trigger on edge rates. Select limits for dV, dt, and slope. Select edge limits between 1 ns and 20 ns.
Exclusion Triggering Trigger on intermittent faults by specifying the expected behavior and triggering when that condition is not met
 

Cascade (Sequence) Triggering

Capability Arm on "A" event, then Trigger on "B" event. Or Arm on "A" event, then Qualify on "B" event, and Trigger on "C" event. Or Arm on “A” event, then Qualify on “B” then “C” event, and Trigger on “D” event
Types A, B, C, or D event: Edge, Glitch, Width, Window, Dropout, Interval, Runt, Slew Rate, or Pattern (analog)
Holdoff Holdoff between A and B, B and C, or both is selectable by time or number of events
Reset Reset between A and B, B and C, or both is selectable in time or number of events
 

High Speed Serial Protocol Triggering

Data Rates 100 Mb/s - 2.7 Gb/s, 3.0, 3.125 Gb/s
Pattern Length 80 bits, NRZ or 8b10b
Clock and Data Outputs 400mVp-p (typical) AC coupled
Clock Recovery Jitter 2 ps rms + 0.3% Unit Interval rms for PRBS data patterns with 50% transition density (typical)
Hardware Clock Recovery Loop BW PLL Loop BW = Fbaud/5500, 100 Mb/s to 2.488 Gb/s (typical)
 

Low Speed Serial Protocol Triggering (Optional)

Optionally available I2C, SPI (SPI, SSPI, SIOP), UART-RS232, CAN, LIN, FlexRay
 

Color Waveform Display

Type Color 15.3" flat panel TFT-Active Matrix LCD with high resolution touch screen
Resolution WXGA; 1280 x 768 pixels.
Number of traces Display a maximum of 8 traces. Simultaneously display channel, zoom, memory and math traces.
Grid Styles Auto, Single, Dual, Quad, Octal, X-Y, Single+X-Y, Dual+X-Y
Waveform Representation Sample dots joined, or sample dots only
 

Internal Waveform Memory

Internal Waveform Memory 4 active waveform memory traces (M1-M4) store 16 bit/point full length waveforms. Waveforms can be stored to any number of files limited only by the data storage media capacity.
 

LeCroy WaveStream™ Fast Viewing Mode

Intensity 256 Intensity Levels, 1–100% adjustable via front panel control
Types Select analog or color-graded
Number of Channels up to 4 simultaneously
Max Sampling Rate 40 GS/s (80 GS/s with optional WM8Zi-2X80GS external interleaving device)
Persistence Aging Select from 500 ms to Infinity
Waveforms/second (continuous) up to 2500 Waveforms/second
Operation Front panel toggle between WaveStream ON (Analog), ON (Color) and OFF
 

Automatic Setup

Auto Setup Automatically sets timebase, trigger, and sensitivity to display a wide range of repetitive signals
Find Vertical Scale Automatically sets the vertical sensitivity and offset for the selected channel to display a waveform with the maximum dynamic range
 

Probes

Probes Qty. (4) ÷10 Passive Probes
Probing System Probus and ProLink. Automatically detects and supports a variety of compatible probes
Scale Factors Automatically or manually selected depending on probe used
Calibration Output 1kHz square wave, 1Vp-p (typical), output to probe hook
 

Integrated Second Display

Type Color 15.3" flat panel TFT-Active Matrix LCD with high resolution touch screen
Resolution WXGA; 1280 x 768 pixels
 

Analog Persistence Display

Analog and Color-Graded Persistence Variable saturation levels; stores each trace’s persistence data in memory
Trace Selection Activate persistence on all or any combination of traces
Sweep Display Modes All accumulated, or all accumulated with last trace highlighted
Persistence Types Select analog, color, or three-dimensional
Persistence Aging Select from 500 ms to infinity
 

High Speed Digitizer Output (Option)

Type LeCroy LSIB
Transfer Rate up to 325 MB/s (typical)
Output Protocol PCI Express, Gen 1 (4 lanes utilized for data transfer)
Control Protocol TCP/IP
Command Set Via Windows Automation, or via LeCroy Remote Command Set
 

Processor/CPU

Type Intel® CoreTM 2 Quad, 2.5 GHz (or better)
Processor Memory 4 GB standard, up to 8 GB optional
(8 GB standard with "M-64", "L-128", or "VL-256" memory)
Operating System Microsoft Windows® Vista Business Edition (64-bit) with SP1
Real Time Clock Date and time displayed with waveform in hardcopy files. SNTP support to synchronize to precision internal clocks.
 

Zoom Expansion Traces

Zoom Expansion Traces Display up to 4 Zoom and 8 Math/Zoom traces
 

Setup Storage

Front Panel and Instrument Status Store to the internal hard drive, over a network, or to a USB-connected peripheral device.
 

Interface

Remote Control Via Windows Automation, or via LeCroy Remote Command Set
Network Communication Standard VXI-11 or VICP, LXI Class C (v1.2) Compliant
GPIB Supports IEEE – 488.2
LSIB Port (optional) Supports PCIe Gen1 x4 protocol with LeCroy supplied API
Ethernet (LAN) Supports 10/100/1000BaseT Ethernet interface (RJ45 port)
USB (Host) minimum 6 total (incl. 3 front panel) USB 2.0 ports support Windows compatible devices
External Monitor Port 15 pin D-Type WXGA compatible to support customer-supplied external monitor.
DVI connector to support LeCroy WPZi-EXTDISP-15 additional touch screen display accessory.
Includes support for extended desktop operation with optional LeCroy or other second monitor.
Peripheral Bus LeCroy LBUS standard
 

Auxiliary Input

Signal Types Select External Trigger or External Clock Input on the front panel
Coupling 50 Ω: DC; 1 MΩ: AC, DC, GND
Max. Input Voltage 50 Ω: 5 Vrms; 1 MΩ: 250 V (Peak AC < 10 kHz + DC)
 

Auxiliary Output

Connector Type BNC, located on front
Signal Types Select from calibrator, control signals or Off
Calibrator Signal 500 Hz–5 MHz square wave or DC level; 0.0 to 500 mV into 50 Ω (0–1 V into 1 MΩ)
Control Signals Trigger enabled, trigger out, pass/fail status, off
 

General

Auto Calibration Ensures specified DC and timing accuracy is maintained for 1 year minimum.
 

Power Requirements

Voltage 100–240 VAC ±10% at 45-66 Hz; 100-120 VAC ±10% at 380-420 Hz; Automatic AC Voltage Selection
Max. Power Consumption 975 W / 975 VA
 

Environmental and Safety

Temperature (Operating) +5 °C to +40 °C including CD-RW/DVD-ROM drive
Temperature (Non-Operating) –20 °C to +60 °C
Humidity (Operating) 5% to 80% relative humidity (non-condensing) up to +31 °C. Upper limit derates to 50% relative humidity (non-condensing) at +40 °C.
Humidity (Non-Operating) 5% to 95% relative humidity (non-condensing) as tested per MIL-PRF-28800F
Altitude (Operating) Up to 10,000 ft. (3048 m) at or below +25 °C
Altitude (Non-Operating) Up to 40,000 ft. (12,192 m)
Random Vibration (Operating) 0.5 grms 5 Hz to 500 Hz, 15 minutes in each of three orthogonal axes
Random Vibration (Non-Operating) 2.4 grms 5 Hz to 500 Hz, 15 minutes in each of three orthogonal axes
Functional Shock 20 g peak, half sine, 11 ms pulse, 3 shocks (positive and negative) in each of three orthogonal axes, 18 shocks total
 

Physical Dimensions

Dimensions (HWD) 14"H x 18.4"W x 15.3"D (355 x 467 x 390 mm)
Weight 51.5 lbs. (23.4 kg)
Shipping Weight 70.0 lbs. (31.8 kg)
 

Certifications

Certifications CE Compliant, UL and cUL listed; conforms to EN 61326, EN 61010-1, UL 61010-1 2nd edition, and CSA C22.2 No. 61010-1-04
 

Warranty and Service

Warranty and Service 3-year warranty; calibration recommended annually. Optional service programs include extended warranty, upgrades, and calibration services.
 

Memory Options

WM8Zi-S-32
32 Mpts/Ch Memory Option for WaveMaster 8 Zi. SDA 8 Zi, and DDA 8 Zi. Includes 4 GB RAM standard
WM8Zi-M-64
64 Mpts/Ch Memory Option for WaveMaster 8 Zi. SDA 8 Zi, and DDA 8 Zi. Includes an additional 4 GB of RAM (8 GB total)
WM8Zi-L-128
128 Mpts/Ch Memory Option for WaveMaster 8 Zi. SDA 8 Zi, and DDA 8 Zi. Includes an additional 4 GB of RAM (8 GB total)
WM8Zi-VL-256
256 Mpts/Ch Memory Option for WaveMaster 8 Zi. SDA 8 Zi, and DDA 8 Zi. Includes an additional 4 GB of RAM (8 GB total)

Sampling Rate Options

WM8Zi-2X80GS
80 GS/s on 2 Ch Sampling Rate Option for WaveMaster 8 Zi (not available for 820Zi, 825Zi or 830Zi).Includes two separate external interleaving devices with storage case.

Computer Upgrade

GPIB-2
GPIB Option for LeCroy oscilloscope. Half-height card with custom connector.
WM8Zi-4-UPG-8GBRAM
Upgrade from 4 GB to 8 GB CPU RAM
WM8Zi-500GB-HD
Upgrade from Standard Size Hard Drive to 500 GB Hard Drive
WM8Zi-160GB-RHD-02
Additional 160 GB Hard Drive. Includes Windows Vista OS, LeCroy Oscilloscope Software and Critical Scope Operational File Duplicates
WM8Zi-500GB-RHD-02
Additional 500 GB Hard Drive. Includes Windows Vista OS, LeCroy Oscilloscope Software and Critical Scope Operational File Duplicates

Serial Data Analysis

WM8Zi-SDAII
SDA II Serial Data Analysis Option WM8Zi-SDAII (Standard on SDA 8 Zi and DDA 8 Zi)
WM8Zi-CBL-DE-EMBED
Cable De-Embed Option (Standard on SDA 8 Zi and DDA 8 Zi)
WM8Zi-EYEDRII
Eye Doctor II Advanced Signal Integrity Tools

Synchronize Oscilloscope Options

Zi-8CH-SYNCH
HW Kit to allow combination of channels from two 8Zi oscilloscopes

Serial Data Compliance

QPHY-DDR2
QualiPHY DDR2 Compliance Software Option
QPHY-DDR3
QualiPHY Enabled DDR3 Software Option
QPHY-DisplayPort
QualiPHY Enabled DisplayPort Software Option
QPHY-ENET
QualiPHY Ethernet Application Software
QPHY-HDMI
QualiPHY Enabled HDMI Software Option (TF-HDMI-3.3V-QUADPAK required;WavePro 760Zi requires SDA II option)
QPHY-PCIe
QualiPHY Enabled PCIe Software Option (for Gen1 and Gen2 testing)
QPHY-SAS2
QualiPHY SAS2 Compliance Software
QPHY-SATA
QualiPHY Enabled SATA Software Option
QPHY-USB
USB Application Software Package
QPHY-USB3-Tx-Rx
QualiPHY Enabled USB 3.0 Software Option
QPHY-UWB
WiMedia UWB Transmitter Measurement Software Option
ENET-2ADA-BNCSMA
2 x BNC to SMA Adapter
ENET-2CAB-SMA018
2 x 18 inch SMA to SMA Cable
ENET-2CAB-SMA036
2 x 36 inch SMA to SMA Cable
WM8Zi-ET-PMT
Electrical Telecom Mask Test Software Package
WM8Zi-SDM
Serial Data Mask Software Package (Standard on SDA 8 Zi and DDA 8 Zi)
TF-DSQ
Probe Deskew and Calibration Fixture
TF-ENET-B
10/100/1000Base-T Compliance Test Fixture
TF-ET
Telecom Adaptor Kit 100Ω, 120Ω, 75Ω
TF-HDMI-3.3V-QUADPAK
HDMI Pull-Up Terminator Quad Pack – for Use with the Efficere ET-HDMI-TPS-S Test Fixture
TF-SATA-C
SATA Compliance Test Fixture (*WavePro 7 Zi and WaveMaster 8 Zi class oscilloscopes require the SDA II software option; **20M memory required)
TF-SATA-C-Kit
SATA Compliance Test Kit (*WavePro 7 Zi and WaveMaster 8 Zi class oscilloscopes require the SDA II software option; **20M memory required)
TF-USB-B
USB 2.0 Testing Compliance Test Fixture
TF-USB3
USB 3.0 Test Fixture Set TF-USB3 ("A" male, "A" female, "B" female)

Serial Trigger and Decode

WM8Zi-HSPT
50 Mb/s to 3.125 Gb/s High-speed Serial Pattern Trigger Option for 4–30 GHz Oscilloscopes and Disk Drive Analyzers (Standard on SDA 8 Zi)
WM8Zi-8B10B D
8b10b Decode Option Standard on SDA 8 Zi and DDA 8 Zi)
WM8Zi-I2Cbus TD
I2C Bus Trigger and Decode Option
WM8Zi-SPIbus TD
SPI Bus Trigger and Decode Option
WM8Zi-LINbus TD
LIN Trigger and Decode Option
WM8Zi-UART-RS232bus TD
UART and RS-232 Trigger and Decode Option
WM8Zi-FlexRayBus TD
FlexRay Trigger and Decode Option
WM8Zi-FlexRayBus TDP
FlexRay Bus Trigger, Decode, and Physical Layer Test Option
WM8Zi-CANbus TDM
CANbus TDM Trigger, Decode and Measure/Graph Option
WM8Zi-CANbus TD
CANbus TD Trigger and Decode Option
WM8Zi-PCIEbus D
PCI Express (Gen 1.x, 2.0, and 3.0) Protocol Link Layer Decode Annotation for WaveMaster 8 Zi Series Oscilloscopes and Analyzers
WM8Zi-AudioBus TD
AudioBus Trigger and decode Option for WaveMaster 8Zi 7Zi
WM8Zi-AudioBus TDG
AudioBus Trigger, decode, and Graph Option for WaveMaster 7Zi
WM8Zi-ProtoSync PE
Oscilloscope & PCIe Protocol Analyzer Synchronization Option
WM8Zi-ProtoSync PE-B
Oscilloscope & PCIe Protocol Analyzer+Bit Tracer Synchronization Option
WM8Zi-ProtoSync PE-B-ExtDisp
Oscilloscope & PCIe Protocol Analyzer+Bit Tracer Synchronization Option with Zi-EXTDISP-15 Display
WM8Zi-1553 TD
MIL-STD-1553 decode option for WaveMaster 8 Zi

High Speed Output Accessories

LSIB-1
High speed PCIe Gen1 x4 digitizer output for LeCroy oscilloscope. Half-height card with custom connector
LSIB-HOSTBOARD
PCI Express x4 Host Interface Board for desktop PC
LSIB-HOSTCARD
PCI Express x1 Express Card Host Interface for laptop Express Card slot
LSIB-CABLE-3M
PCI Express x 4 3-meter cable with x4 cable connectors included
LSIB-CABLE-7M
PCI Express x 4 7-meter cable with x4 cable connectors include

Mixed Signal Solutions

MS-500
500 MHz Maximum Input Frequency, 18 Channels, 50 Mpts/ch Mixed Signal Oscilloscope Option
MS-500-36
250 MHz Maximum Input Frequency, 36 Channels, 25 Mpts/ch Mixed Signal Oscilloscope Option
MS-250
250 MHz Maximum Input Frequency, 18 Channels, 10 Mpts/ch Mixed Signal Oscilloscope Option

Mixed Signal Accessories

PK400-1
Large gripper probe set for 0.10'' (2.54 mm) pin pitch, includes 10 probes with color-coded leads
PK400-2
Medium gripper probe set for 0.04'' (1.0 mm) pin pitch, includes 10 probes with color-coded leads
PK400-3
Small gripper probe set for 0.008'' (0.2 mm) pin pitch, includes 10 probes with color-coded leads
MSO-MICTOR
Mictor Cable, D0-D35
MSO-3M
3M Interconnect Cable, D0-D17 or D18-D35

Jitter Analysis Software

WM8Zi-JTA2
Jitter Timing and Analysis Software Package (Standard on SDA8 Zi and DDA 8 Zi)

Other Software Options

WM8Zi-XDEV
Advanced Customization Software Package
WM8Zi-SPECTRUM
Spectrum Analyzer and Advanced FFT Option
WM8Zi-EMC
EMC Pulse Parameter Software Package

Data Storage Software

WM8Zi-AORM
Advanced Optical Recording Measurement Package
WM8Zi-DDM2
Disk Drive Measurements Software Package (Standard on DDA 8 Zi)
WM8Zi-DDA
Disk Drive Analyzer Software Package

Power Analysis Software

WM8Zi-PMA2
Power Measure Analysis Software Package

Digital Filtering Software

WM8Zi-DFP2
Digital Filter Software Package

Miscellaneous

OC1024
OC1024 scope cart is sturdy and mobile
OC1021
OC1021 scope cart is sturdy and mobile
KYBD-1
Keyboard
ZI-EXTDISP-15
Top-mounted, Fully Integrated 15.3" WXGA with Touch Screen Display, Including all Cabling and Software
WM8Zi-HARDCASE
Hard Carrying Case
WM8Zi-SOFTCASE
Soft Carrying Case
WM8Zi-RACKMOUNT
Rackmount Accessory for Converting aWM8Zi Series Oscilloscope to an 8U Rack-mounted Package
TTL-AUX-OUT
Accessory for Zi oscilloscopes to enable TTL level output from the Aux Out connector
Zi-FrontPanel
Removable front panel for WavePro 7Zi and WaveMaster 8Zi Oscilloscopes
Zi-FrontPanel-4Ch
Removable front panel with 4 independent channel controls for WavePro 7Zi and WaveMaster 8Zi Oscilloscopes

Active Voltage Probes

ZS1500-QUADPAK
Set of 4 ZS1500 Active Probes
ZS1000-QUADPAK
Set of 4 ZS1000 Active Probes
ZS1500
1.5 GHz, 1 MΩ, 0.9 pF Active Probe
ZS1000
1 GHz, 1 MΩ, 0.9 pF Active Probe
HFP2500
2.5 GHz Active Voltage Probe

Current Probes

CP030
30 Amps, 50 MHz BW Current Probe - AC/DC , 50 Amp Peak Pulse
CP031
30 Amps, 100 MHz BW Current Probe - AC/DC , 50 Amp Peak Pulse
AP015
30 Amps, 50MHz BW Current Probe - AC/DC, 50 Amp Peak Pulse
CP150
150 Amps, 10 MHz BW Current Probe - AC/DC , 500 Amp Peak Pulse
CP500
500 Amps, 2 MHz BW Current Probe - AC/DC , 700 Amp Peak
DCS015
Deskew Calibration Source for AP015

Differential Amplifiers

DA1855A
DA1855A, 100 MHz Differential Amplifier with fast over drive recovery, calibrated offset, and selectable LP filters
DXC100A
DXC100A, Selectable ÷100 or ÷10 , Passive Differential Probe Pair for DA18xxA series
DXC200
÷1,50 MHz low capacitance Passive Differential Probe Pair for DA18xxA series
DXC-5100
÷100 2.5KV Passive High Voltage Probe Pair for DA18xxA series. Requires DA101 for full performance
DA101
÷10 1MOhm Passive Attenuator for DXC series probes
DA1855A-PR2
2 Ch, 100 MHz Differential Amplifier with fast over drive recovery, calibrated offset, and selectable LP filters
DA1855A-PR2-RM
Rack mountable version of the DA1855A-PR2
DA1855A-RM
DA1855A with rackmount

Differential Probes

D2505
WaveLink D2505 25 GHz/1.6Vp-p Differential Probe with Dxx05-SI Solder-In Tip (Qty. 2).
D2505-PS
25 GHz Complete Probe System with Solder-In Tip (25 GHz) and Positioner Tip Browser (22 GHz)
D2005
WaveLink D2005 20 GHz/1.6Vp-p Differential Probe with Dxx05-SI Solder-In Tip (Qty. 2).
D2005-PS
20 GHz Complete Probe System with Solder-In Tip (20 GHz) and Positioner Tip Browser (20 GHz)
D1605
WaveLink D1605 16 GHz/1.6Vp-p Differential Probe with Dxx05-SI Solder-In Tip (Qty. 2).
D1605-PS
16 GHz Complete Probe System with Solder-In Tip (16 GHz) and Positioner Tip Browser (16 GHz)
D1305
WaveLink D1305 13 GHz/1.6Vp-p Differential Probe with Dxx05-SI Solder-In Tip (Qty. 2).
D1305-PS
13 GHz Complete Probe System with Solder-In Tip (13 GHz) and Positioner Tip Browser (13 GHz)
Dxx05-PT-KIT
WaveLink Dxx05-PT (22 GHz rating) Adjustable Positioner Tip Kit. For use with Dxx05 amplifiers.
WL-PLink-A
WaveLink ProLink Platform/Cable Assembly Kit for >=13 GHz WaveLink Probes.
LPA-2.92
ProLink to 2.92mm Adapter with Probe Power and Communication Pass Through
D620
WaveLink 6 GHz Differential probe module 5V
D610
WaveLink 6 GHz Differential probe module 2.5V
D320
WaveLink 3.5 GHz Differential probe module 5V
AP034
1 GHz Active Differential Probe
D310
WaveLink 3.5 GHz Differential probe module 2.5V
Dx10-PT-KIT
WaveLink Dx10-PT Adjustable Positioner Tip Kit. Includes XYZ Positioner Assembly with Mechanical Interconnects, Adhesive Kit, Connection Guides, Hand-held Wand, and Replacement Pins(Qty. 4)/Sockets(Qty 2).
Dx20-PT-KIT
WaveLink Dx20-PT Adjustable Positioner Tip Kit. Includes XYZ Positioner Assembly with Mechanical Interconnects, Adhesive Kit, Connection Guides, Hand-held Wand, and Replacement Pins (Qty. 4)/Sockets(Qty. 2).
D600A-AT
WaveLink 7.5 GHz Differential Probe with Adjustable Tip Module
D500PT
WaveLink 6 GHz Differential Positioner Mounted Tip Module
WL-PLink
WaveLink ProLink probe body >3.5 GHz
WL-Pbus
WaveLink ProBus probe body = 3.5 GHz
AP033
500 MHz Active Differential Probe
DA18000AC
18 GHz Type “K” Differential Probe, AC Coupled
D13000PS
13 GHz Differential Probe System
D11000PS
11 GHz Differential Probe System

High Voltage Probes

PPE1.2KV
10:1/100:1 200/300 MHz 5M Ohm/50M Ohm
PPE2KV
100:1 400 MHz 50M Ohm High Voltage Probe
PPE4KV
100:1 400 MHz 50M Ohm High-Voltage Probe
PPE5KV
100:1 400 MHz 50M Ohm High-Voltage Probe
PPE6KV
1000:1 400 MHz 50M Ohm High-Voltage Probe
PPE20KV
1000:1 100 MHz 100M Ohm High-Voltage Probe

High Voltage Differential Probes

AP031
15 MHz Differential Probe
ADP300
High-Voltage Active Differential Probe
ADP305
High-Voltage Differential Probes From LeCroy

Optical Probes

OE425
Short wavelength O/E Converter (500-870 nm)
OE455
Long wavelength O/E Converter (950-1630 nm)
OE525
Short wavelength O/E Converter (500-870 nm)
OE555
Long wavelength O/E converter

Probe Parts and Kits

LPA-SMA-KIT-A
Kit of Pro-Link to SMA Adapter, new style
LPA-SMA-A
Pro-Link to SMA Adapter, new style
LPA-K-KIT-A
Kit of ProLink to K/2.92 mm Adapters
LPA-K-A
ProLink to K/2.92 mm Adapter
DXX05-SI
Replacement Dxx05-SI 13-25 GHz Solder-In Lead with Qty. 10 Spare Resistors. For warranted accuracy, amplifier must be returned to factory for calibration with this lead.
DXX05-PT
Replacement Dxx05-PT Positioner Tip
Dx20-SP
Replacement Square Pin lead set for D320/D620
Dx20-SI
Replacement Solder tip lead set for D320/D620
Dx20-QC
Replacement Quick Connect lead set for D320/D620
Dx20-PT
Replacement Dx20-PT Tip with Qty. 4 Pogo Pin Tips and Qty. 2 Sockets
Dx10-SP
Replacement Square Pin lead set for D310/D610
Dx10-SI
Replacement Solder tip lead set for D310/D610
Dx10-PT
Replacement Dx10-PT Tip with Qty. 4 Pogo Pin Tips and Qty. 2 Sockets
D620-PT
WaveLink D620 6 GHz/5Vp-p and +/-3V Offset Range Differential Amplifier Small Tip Module with Dx20-PT Adjustable Positioner Tip. 4.8 mVrms (typical) probe + scope noise. Includes XYZ Positioner Assembly with Mechanical Interconnects, Adhesive Kit, Connection Guides, Hand-held Wand, and Replacement Pins (Qty. 4)/Sockets(Qty. 2).
D610-PT
WaveLink D610 6 GHz/2.5Vp-p and +/-3V Offset Range Differential Amplifier Small Tip Module with Dx10-PT Adjustable Positioner Tip. 2.8 mVrms (typical) probe + scope noise. Includes XYZ Positioner Assembly with Mechanical Interconnects, Adhesive Kit, Connection Guides, Hand-held Wand, and Replacement Pins (Qty. 4)/Sockets(Qty. 2).
D600ST
WaveLink 7 GHz Differential Probe with Small Tip Module (includes D6000ST-SI, D600ST-QC, D600ST-SP)
D350ST
WaveLink 4 GHz, 5 V Differential with Small Tip Module
D320-PT
WaveLink D320 3.5 GHz/5Vp-p and +/-3V Offset Range Differential Amplifier Small Tip Module with Dx20-PT Adjustable Positioner Tip. 4.8 mVrms (typical) probe + scope noise. Includes XYZ Positioner Assembly with Mechanical Interconnects, Adhesive Kit, Connection Guides, Hand-held Wand, and Replacement Pins (Qty. 4)/Sockets(Qty. 2).
D310-PT
WaveLink D310 3.5 GHz/2.5Vp-p and +/-3V Offset Range Differential Amplifier Small Tip Module with Dx10-PT Adjustable Positioner Tip. 2.8 mVrms (typical) probe + scope noise. Includes XYZ Positioner Assembly with Mechanical Interconnects, Adhesive Kit, Connection Guides, Hand-held Wand, and Replacement Pins (Qty. 4)/Sockets(Qty. 2).
D300A-AT
WaveLink 4 GHz Differential Probe with Adjustable Tip Module
D13000SI
Replacement Solder-In Interconnect Lead
D11000SI
Replacement solder-in interconnect lead

Transmission Line Probes

PP066
7.5 GHz Low Capacitance ÷10/÷20 probe