WaveJet 332A
350 MHz, 1 GS/s, 2 Ch, 500 kpts/Ch with 7.5" Color Display. 2 GS/s Interleaved
WaveJet 312A (100 MHz Bandwidth, 2 Input Channels, 1 GS/s Max Sample Rate) WaveJet 314A (100 MHz Bandwidth, 4 Input Channels, 1 GS/s Max Sample Rate) WaveJet 322A (200 MHz Bandwidth, 2 Input Channels, 1 GS/s (all channels), 2 GS/s (Interleaved) Max Sample Rate) WaveJet 324A (200 MHz Bandwidth, 4 Input Channels, 1 GS/s (all channels), 2 GS/s (Interleaved) Max Sample Rate) WaveJet 332A (350 MHz Bandwidth, 2 Input Channels, 1 GS/s (all channels), 2 GS/s (Interleaved) Max Sample Rate) WaveJet 334A (350 MHz Bandwidth, 4 Input Channels, 1 GS/s (all channels), 2 GS/s (Interleaved) Max Sample Rate) WaveJet 352A (500 MHz Bandwidth, 2 Input Channels, 1 GS/s (all channels), 2 GS/s (Interleaved) Max Sample Rate) WaveJet 354A (500 MHz Bandwidth, 4 Input Channels, 1 GS/s (all channels), 2 GS/s (Interleaved) Max Sample Rate)
Long Capture Time The 500 kpts/Ch memory eliminates the tradeoff between long capture and high sample rate providing long capture at full 2 GS/s sample rate. The long memory makes the WaveJet a great tool for viewing low frequency and high frequency signals or signals with fast edges. 
Waveform Math The WaveJet provides math capabilities for additional analysis. Available math functions include sum, difference, product and FFT. Measurements can then be made on the calculated waveforms using the parameters or cursors to provide additional debug and analysis capabilities. 
Automatic Measurements Save time making measurements on your signals by using the 26 automatic measurement parameters. See your results color coded to the channels that are being measured. For a more in-depth look turn on the min/max statistics to observe trends in the measurements. 
Replay Mode The fast update rate shows runts and glitches when they occur but it is hard to tell exactly when they occurred. Replay mode lets you go back in time to isolate those anomalies, measure them with parameters or cursors, and quickly find the source of the problem. 
Advanced Triggering Along with edge triggering, additional triggering capabilities include Pulse Width, Period, Pulse Count and TV to help you capture the signals you need to see. 
Frequency Counter Use the built in 6-digit frequency counter to simplify how you make measurements. The counter is always displayed and easy to read. Acquisition Modes Peak detect and equivalent time acquisition modes offer flexibility in how you capture and measure your signals. The WaveJet can capture glitches as small as 1 ns with peak detect and can achieve a sampling rate of up to100 GS/s with equivalent time sampling. Connectivity and Communication Saving waveforms and screen images is an important part of documenting results. The WaveJet has a front panel USB port to save data to memory stick and a rear panel USB for printing hardcopies. 
The rear panel USB port, along with optional GPIB and Ethernet connections provide full remote control of the instrument. LeCroy's Scope Explorer and ActiveDSO software utilities provide a quick method to begin controlling the WaveJet.
Vertical System | | Bandwidth | 350 MHz | | Rise time | 1 ns | | Input Channels | 2 | | Bandwidth Limiters | 20 MHz, 100 MHz | | Input Impedance | 1 MΩ ± 1.5% || 16 pF, 50 Ω ± 1.5% | | Input Coupling | AC, DC, GND | | Maximum Input Voltage | 1 MΩ: ± 400 Vpk, 50 Ω: 5 VRMS | | Channel-Channel Isolation | ≥ 34 dB from DC - 100 MHz ≥ 30 dB at 350 MHz | | Vertical Resolution | 8 bits | | Sensitivity | 1 MΩ: 2 mV/div - 10 V/div 50 Ω: 2 mV/div - 2 V/div | | DC Gain Accuracy | ±(1.5% + 0.5% of Full Scale) | | Offset Range | ±1 V: 2 mV/div - 50 mV/div ±10 V: 50.2 mV/div - 500 mV/div ±100 V: 502 mV/div - 10 V/div | | | Horizontal System | | Time/Division Range | 1 ns/div - 50 s/div | | Clock Accuracy | 10 ppm | | Trigger Jitter | 200 ps (pk-pk) | | | Acquisition System | | Single-Shot Sample Rate/Ch | 2 GS/s on 1 Ch 1 GS/s on 2 Ch | | Equivalent Sample Rate | 100 GS/s | | Memory | 500 kpts/Ch | | | Acquisition Modes | | Averaging | Selectable Number of Sweeps:
2, 4, 16, 32, 64, 128, 256 | | Peak Detect | 1 ns | | Interpolation | Linear | | | Triggering System | | Modes | Normal, Auto, Single, Stop | | Sources | Ch 1 - Ch 2, EXT, EXT/10, AC Line | | Coupling Mode | AC, DC, LF Rej, HF Rej, | | Hold-off by Time | 200 ns - 50 s | | Internal Trigger Range | ±5 divisions from center | | Trigger Sensitivity | 0.5 division: DC - 10 MHz 2.0 divisions: 10 MHz - 350 MHz | | External Trigger Input Range | Ext: ±0.5 V, EXT/10: ±5.0 V | | | Display | | Type | Color 7.5" TFT-LCD | | Resolution | VGA: 640 x 480 | | Grid Styles | YT, XY | | | Internal Storage | | Waveform Storage | 1 reference waveform | | Setup Storage | 5 setups | | | Probes | | Probes
| Qty. (2) ÷10 Passive Probes | | Scale Factors | Automatic: ÷1, ÷10, ÷100, ÷1000 Manual: ÷1, ÷10, ÷20, ÷100, ÷200, ÷1000, ÷2000 | | Calibration Output | 1 kHz square wave, 0.6 Vp-p (typical), output to probe hook | | | Interface | | GPIB | 1 (optional) | | USB (Device) | 1 | | Ethernet (LAN) | 1 (optional) | | USB (Host) | 1 | | Serial (RS-232) | na | | | Power Requirements | | Voltage | 90 - 264 VAC, 47 HZ - 420 Hz | | Max. Power Consumption | 75 VA | | | Environmental and Safety | | Temperature (Operating) | 0 °C to 40 °C | | Temperature (Non-Operating) | -20 °C to 60 °C | | Humidity (Operating) | 80% RH, 40 °C | | Humidity (Non-Operating) | 80% RH, 40 °C | | Altitude (Operating) | 2000 m | | Altitude (Non-Operating) | 15000 m | | | Physical Dimensions | | Dimensions (HWD) | 190 mm x 285 mm x 102 mm (7.5" x 11.2" x 4") | | Weight | 3.2 kg (7 lbs) | | |
LogicStudio | LogicStudio 16 16 Channel, 1 GS/s, 100 MHz USB Logic Analyzer | Remote Control/Network Options | WJ-A-LAN 10/100Base-T Interface for WaveJet 300A Series | WJ-A-GPIB GPIB Interface for WaveJet 300A Series | Miscellaneous | WJ-CASE WaveJet Carrying Case and Protective Front Cover | WJ-RACK 5U Rackmount Kit for WaveJet |
High Voltage Probes | PPE1.2KV 10:1/100:1 200/300 MHz 5M Ohm/50M Ohm | PPE2KV 100:1 400 MHz 50M Ohm High Voltage Probe | PPE4KV 100:1 400 MHz 50M Ohm High-Voltage Probe | PPE5KV 100:1 400 MHz 50M Ohm High-Voltage Probe | PPE6KV 1000:1 400 MHz 50M Ohm High-Voltage Probe | PPE20KV 1000:1 100 MHz 100M Ohm High-Voltage Probe | High Voltage Differential Probes | AP031 15 MHz Differential Probe |
| Product | Price* | Description | | | WaveJet 332A | $5,090.00 | 350 MHz, 1 GS/s, 2 Ch, 500 kpts/Ch with 7.5" Color Display. 2 GS/s Interleaved | Request Quote | | WaveJet 312A | $3,050.00 | 100 MHz, 1 GS/s, 2 Ch, 500 kpts/Ch with 7.5" Color Display | Request Quote | | WaveJet 314A | $3,560.00 | 100 MHz, 1 GS/s, 4 Ch, 500 kpts/Ch with 7.5" Color Display | Request Quote | | WaveJet 322A | $3,815.00 | 200 MHz, 1 GS/s, 2 Ch, 500 kpts/Ch with 7.5" Color Display. 2 GS/s Interleaved | Request Quote | | WaveJet 324A | $4,580.00 | 200 MHz, 1 GS/s, 4 Ch, 500 kpts/Ch with 7.5" Color Display. 2 GS/s Interleaved | Request Quote | | WaveJet 334A | $5,855.00 | 350 MHz, 1 GS/s, 4 Ch, 500 kpts/Ch with 7.5" Color Display. 2 GS/s Interleaved | Request Quote | | WaveJet 352A | $6,110.00 | 500 MHz, 1 GS/s, 2 Ch, 500 kpts/Ch with 7.5" Color Display. 2 GS/s Interleaved | Request Quote | | WaveJet 354A | $6,930.00 | 500 MHz, 1 GS/s, 4 Ch, 500 kpts/Ch with 7.5" Color Display. 2 GS/s Interleaved | Request Quote |
* Prices Shown Are US Domestic Prices, MSRP, and are exclusive of local and state taxes, shipping, levies, fees, duties, exportation/importation costs, service and warranties outside the United States, and assume standard 30 day payment terms. Information and/or pricing may be changed or updated without notice. LeCroy may also make improvements and/or changes in the products and/or the programs, as well as associated pricing, described in this information at any time without notice. Because international information is provided at this Web Site, not all products, programs or pricing mentioned will be available in your country. Please contact your local sales representative for information as to products and services available in your country.
|
|
|