Customize
Your Scope
CustomDSO
Files Create Custom Menus and Operations
CustomDSO
files allow users to create custom menus with a series of user specific
measurement operations behind them. Figure 1 includes a typical use of the CustomDSO file.
Each user created menu in this example sets up a series of specific tests
including both mask and parameter testing on multiple waveforms.
An experienced scope users can easily produce a semi-automated test
sequence which permits users with limited scope familiarity to operate the most
advanced features of the scope.

Figure
1 -
Example of a CustomDSO based test where the DSO menus setup specific tests
involving mask and parameter testing on multiple waveforms
CustomDSO is
based on ASCII text files labeled with a .DSO extension.
Each of the .DSO files contains a series of oscilloscope remote commands.
These files are input to the scope using a floppy disk or SRAM memory
card. At power-on, the scope looks
for the primary file, named 'AUTOEXEC.DSO' located in the LECROY_1.DIR
directory, and executes the commands found in the file.
The commands in the primary file can call additional .DSO files so that
relatively complex measurement sequences can be implemented.
The CustomDSO
files are created and edited in a text editor such as the Windows Notepad. Figure
2 shows a typical AUTOEXEC.DSO file.

Figure
2 - A
typical autoexec.dso file showing the initial labeling of the menu keys (using
the KEY command) and the destination .dso files for each selection
The most
frequently used commands in CustomDSO files are:
EKEY
- Enable key labels
KEY
n - Labels menu key (1-5) and
defines the .DSO file to
run if that key is pressed.
MESSAGE
- Prints a message on
the scope screen
The other
remote commands used in the .DSO files can be found in the scope's Remote
Control Manual. An easier way to
obtain the more complex commands is to set up the scope function you are
interested in and use ScopeExplorer (a free utility) to query the command.
Cut and paste the response from ScopeExplorer into the .DSO file.
Figure 3 shows an example of recalling the command for setting up trace A
to perform the FFT of channel 2, which had be set up manually.
The define query in the upper half of the figure reads the scope's
existing setup, which appears in the lower half of the figure as "TA:DEF
EQN,
."

Figure
3 - Using ScopeExplorer to learn the remote commands for a math setup
Another way
to simplify the initial setup of the scope is to set all the controls of the
scope up manually then save the entire state using a panel setup to floppy disk.
The remote command:
"RCPN
DISK,FLPY,"P000.PNL'"
reads the
panel setup from the floppy disk and returns the scope to the desired set up.
The internal
Remote Control Assistant (RCA) is a powerful tool for debugging both remote
operation and internal CustomDSO file operation.
RCA is accessible under the Utilities menu, special modes.
Figure 4 shows the use of RCA for debugging a command trying to turn on a
non-existent trace.

Figure
4 Example
of the Remote Control Assistant display showing an error (trace header, TE,
names a trace that doesn't exist)
CustomDSO
files offer the opportunity to create custom default setups, provide
semi-automated prompted operation, or create short cuts to commonly used
features.
|