9384 Series Digital
Oscilloscope
1 GHz Bandwidth, 1 - 4 GS/s
Main Features
| 1 GHz Bandwidth Sample rates to 4 Gigasamples/second
Memory lengths to
8M points
8-bit vertical resolution,
11 with ERES option
Four full independent channels
Vertical accuracy to 1% typical
Triggering to 1 GHz
Innovative Peak Detect
Glitch, Pattern, Qualified, Interval, Dropout, TV, and Exclusion Triggers
Histogram and FFT Signal Processing Options
Floppy Disk, Internal Printer, PCMCIA Hard Disk and Memory Card available
Fully programmable via GPIB and RS-232-C |
1 GHz Bandwidth
The 9384 series digital storage oscilloscope opens up new horizons for engineers and
scientists at the leading edge of technological developments. With 1 GHz bandwidth
and long acquisition memories, it is now possible to reveal previously hidden waveform
details. Narrow glitches are more accurately defined; risetime measurements below 1
nanosecond are more precise; and high-frequency content, filtered out in lower bandwidth
systems, is retained, thereby preserving signal amplitudes and overall signal integrity.
Edge Triggering to 1GHz and LeCroy's powerful Smart Trigger modes like: Glitch,
Pattern, Dropout, Interval, Exclusion and TV enable you to capture precisely the events of
interest.
4 GS/s Sample Rate
The 9384 samples simultaneously on all channels at 1 GS/s. Thus, it is ideal for
demanding high speed applications. In addition, two channels can be combined to provide a
sample rate of 2 GS/s or 4 GS/s in single channel mode. Finer horizontal resolution
and accuracy are assured by high sample rates. This is especially critical in digital
design where unpredictable circuit behavior needs to be identified and analyzed in detail
to be fully understood. Together with this excellent single-shot performance the 9384
series also provides a sample rate equivalent to 10 GS/s for repetitive signals. The
innovative peak-detect mode enables glitch capture even at the slowest time settings
without loss of precision.
8M Points Acquisition Memory
Channel record lengths of 100k, 500k and 2M are available on the 9384. The memory power
is revealed when the user seeks to sample at the highest speed over many timebase
settings. DSO's with less memory may boast a high sample rate for short waveforms, but
only LeCroy's long memory oscilloscope can deliver high sample rates for long waveforms.
To exploit this capability to its fullest, the LeCroy 9384AL combines its channel
acquisition memories to give the user up to 8 million sample points, thereby providing the
waveform detail required on long and complex signals.
The combined capabilities of the 9384 place it in the forefront of DSO capability.
Precision Acquisition
1 GHz bandwidth results in greater accuracy of amplitude and risetime measurements for
high frequency signals and true representation of high speed digital and analog signals.
High Sample Rate
The 9384 provides maximum sample rates of 1 GS/s, 2 GS/s, and 4 GS/s for greater
waveform fidelity, excellent zoom detail, protection against aliasing, precision time
resolution and wider frequency spectrum.
Channel Memory and Channel Interleaving
Capturing long time windows single-shot requires long memory. LeCroy, the company that
invented long memory DSO's, provides three memory selections: 100k points per channel
standard, 500k points per channel with the M and TM versions, and
2M points per channel with the AL version.
By interleaving the 4 channels of the 9384AL, the acquisition memory and sample rate
can be increased 4X to provide up to 4 GS/s and up to 8M-points of memory. The result is
that the 9384AL enables you to capture 2 millisecond of signal duration with 250 ps real
time sample resolution.
Advanced Peak Detect System
The 9384 series offers an innovative peak detect capture mode. This captures fast
glitches by running the ADC's at a high sampling rate even at slow time base settings
thereby capturing signal details that might have been missed due to under sampling. At the
same time the scope stores the underlying data to ensure no loss of time precision -
unlike other peak detect systems.
SMART Trigger System
SMART Trigger functions including Glitch, Pattern, Interval, Exclusion, TV, Dropout,
and State-or-Edge Qualified triggers are available. Pre and Post-trigger delay are fully
variable, Time and Events Holdoff are also included.
Automatic Parametric Measurements and Statistics
The 9384 provides more than 40 parametric measurements and their Average, Highest,
Lowest values and Standard Deviation. Pass/Fail testing allows up to 5 parameters to be
tested against selectable thresholds. Waveform Limit Testing can also be performed using
Masks which may be defined inside the instrument. Any failure will activate preprogrammed
actions such as Hardcopy, Save, Stop, Beep, GPIB SRQ, or Pulse Out.
Internal Printer
Most printers and plotters can be driven via GPIB, RS-232-C and an optional Centronics
interface. The 9384 offers an optional internal printer which can produce a 126 X 90 mm
full resolution screen dump in under 10 seconds at the push of a button.
The unique `Strip-Chart' format expands the horizontal axis up to 200 cm/div per
division for viewing fine waveform detail within long memory acquisitions.
Remote Interfacing
GPIB and RS-232-C interfaces may be used for full remote control of the instrument. All
front panel and internal processing functions can be controlled via either interface.
Optional Advanced Math Package - WP01
Option WP01 provides Summed and Continuous Averaging, Waveform Math Functions, Extrema
and Enhanced Resolution Modes.
Functions can be chained together, allowing complex computations. Waveform operations
can be performed on live, stored, processed or expanded waveforms without altering the
original capture data.
Optional Spectral Analysis Package - WP02
Option WP02 provides comprehensive Spectral Analysis capabilities permitting the system
designer to identify characteristics which may not be apparent in the time domain. WP02
provides a wide selection of windowing functions as well as averaging in the frequency
domain (see page 59 of Catalog).
Spectral analysis can be performed on repetitive and single-shot waveforms, on any part
or zoomed portions of a waveform up to 4M points, not just the first 10k points of the
waveform.
Optional Parameter Analysis Package - WP03
Option WP03 provides extensive parameter analysis capabilities. Detailed statistical
analysis can easily be performed on difficult to measure waveform phenomena such as
amplitude fluctuation and timing jitter. Live histogram displays represent the statistical
distribution of the user selected waveform parameter measurements. Statistical information
can be extracted directly from the histograms using automatic statistical measurement
including max, min, average, median, and std. deviation (see page 63).
ACQUISITION SYSTEM
Bandwidth (-3 dB):
@ 50 Ohm: DC to 1 GHz 10 mV/div and above
@ 1M Ohm DC: DC to 500 MHz typ. at probe tip, with PP005 supplied standard. 1 GHz FET
probe optional.
No. of Channels: 4 - 9384
No. of Digitizers: 4 - 9384
Maximum Sample Rate: 4 GS/s
Memories: (See table)
Sensitivity: 2 mV/div to 1 V/div, 50 Ohm fully variable, 2 mV/div to 10
V/div, 1M Ohm fully variable.
Scale factors: A wide choice of probe attenuation factors are selectable.
Offset Range:
2.00 - 4.99 mV/div: ± 400 mV
5.00 - 99 mV/div: ± 1 V
0.1 - 1.0 V/div: ± 10 V
1.0 - 10 V/div: ±100 V (1M Ohm only)
± 20 V across the whole sensitivity range when using the AP 020 FET probe.
DC Accuracy: 10 mV and above, 1% typical
Vertical Resolution: 8 bits
Bandwidth Limiter: 25 MHz or 200 MHz user selectable
Input Coupling: AC, DC, GND.
Input Impedance: 1 M Ohm // 15 pF typical or 50 Ohm ±1%
Max Input:
1 M Ohm: 400 V (DC+peak AC £10 kHz)
50 Ohm: ± 5 V DC
| Channels |
Maximum |
Memory Per |
|
|
|
Active |
| Use |
Sample Rate |
Channel |
|
|
|
Channels |
| |
|
|
|
|
|
|
| |
|
9384 |
9384M |
9384TM |
9384AL |
|
| |
|
|
|
|
|
|
| All Peak Detect |
1 GS/s |
100k |
500k |
500k |
2M |
All |
| OFF |
|
|
|
|
|
|
| |
|
|
|
|
|
|
| Paired Peak |
2 GS/s |
200k |
1M |
1M |
4M |
Ch 2 & Ch 3 |
| Detect OFF |
|
|
|
|
|
|
| |
|
|
|
|
|
|
| Paired + PP094 |
4 GS/s |
400k |
2M |
2M |
8M |
(PP094 input) |
| Peak Detect |
|
|
|
|
|
|
| OFF |
|
|
|
|
|
|
| |
|
|
|
|
|
|
| All 2.5 ns Peak |
100 MS/s |
50k data+50k |
250k data+ |
250k data+ |
1M data+ |
All 2.5 ns |
| Detect ON |
data+ |
peak |
250k peak |
250k peak |
1M peak |
Detect |
| |
400 MS/s peak |
|
|
|
|
|
TIME BASE SYSTEM
Timebases: Main and up to 4 Zoom Traces.
Time/Div Range: 1 ns/div to 1000 s/div.
Clock Accuracy: < 10 ppm
Interpolator resolution: 10 ps
Roll Mode: ranges 500 ms to 1,000 s/div.
For > 50k points: 10 s to 1,000 s/div.
TRIGGERING SYSTEM
Trigger Modes: Normal, Auto, Single.
Trigger Sources: CH1, CH2, CH3 and CH4, External and Line, Slope, Level
and Coupling for each source can be set independently.
Slope: Positive, Negative.
Coupling: AC, DC, HF, LFREJ, HFREJ.
Pre-trigger recording: 0 to 100% of full scale (adjustable in 1%
increments).
Post-trigger delay: 0 to 10,000 divisions (adjustable in 0.1 div.
increments).
Holdoff by time: 10 ns to 20 s.
Holdoff by events: 0 to 99,999,999 events.
Width Trigger Sensitivity:
<10% of full scale > 1 ns;
<20% of full scale 500 ps to 1 ns
Internal Trigger Sensitivity Range: ± 5 div.
EXT Trigger Max Input:
1 M Ohm // 15 pF: 400 V (DC+peak AC<10 kHz)
50 Ohm± 1%:
± 5 V DC (500 mW) or 5V RMS
EXT Trigger Range: ± 0.5 V (± 5 V with Ext/10)
Trigger Timing: Trigger Date and Time are listed in the Waveform Status
Menu.
SMART TRIGGER TYPES
Pattern: Trigger on the logic AND of 5 inputs - CH1, CH2, CH3, CH4, and
EXT Trigger, where each source can be defined as High, Low or Dont Care. The Trigger
can be defined as the beginning or end of the specified pattern.
Signal or Pattern Width: Trigger on glitches <2.5 ns (1 ns typical) or
on pulse widths between two limits selectable from <2.5 ns to 20s exclusive.
Exclusion Trigger: Trigger on a signal or period outside two limits
selectable from <2.5 ns to 20s.
Signal or Pattern Interval: Trigger on an interval between two limits
selectable from 10 ns to 20 s.
Dropout: Trigger if the input signal drops out for longer than a time-out
from 25 ns to 20 s.
State/Edge Qualified: Trigger on any source only if a given state (or
transition) has occurred on another source. The delay between these events can be defined
as a number of events on the trigger channel or as a time interval.
TV: Allows selection of both line (up to 1500) and field number (up to 8)
for PAL, SECAM, NTSC or non-standard video.
ACQUISITION MODES
Random Interleaved Sampling (RIS): for repetitive signals from 1 ns/div
to
2 ms/div.
Random Interleaved Sampling Rate:
10 GS/s
Single shot: for transient and repetitive signals from 2 ns/div (all
channels active).
Peak detect: captures and displays 2.5 ns glitches (1 ns typical) or
other high-speed events.
Sequence: Stores multiple events - each of them time stamped - in
segmented acquisition memories.
Number of segments available:
9384 2-500
9384M 2-2,000
9384TM 2-2,000
9384AL 2-2,000
DISPLAY
Waveform style: Vectors connect the individual sample points, which are
highlighted as dots. Vectors may be switched off.
CRT: 12.5 x 17.5 cm (9" diagonal) raster.
Resolution: 810 x 696 points.
Modes: Normal, X-Y, Variable or Infinite Persistence.
Real-time Clock: Date, hours, minutes, seconds.
Graticules: Internally generated; separate intensity control for grids
and waveforms.
Grids: 1, 2 or 4 grids.
Formats: YT, XY, and both together.
Vertical Zoom: Up to 5x Vertical Expansion (25x with averaging, up to 80
µV sensitivity with Advanced Math option WP01).
Horizontal Zoom Factors up to:
9384 20,000x
9384M 100,000x
9384TM 100,000x
9384AL 200,000x
Waveforms can be expanded to give .4-.5 points/division. Zoom factors up to 800,000x for
the 9384L with all channels combined.
INTERNAL MEMORY
Waveform Memory: Up to four 16-bit Memories (M1,M2,M3,M4). The length of
each memory is equal to the data acquisition memory.
Processing Memory: Up to four 16-bit Waveform Processing Memories
(A,B,C,D).
Setup Memory: Four non-volatile memories. Optional IC Memory Cards,
floppy disk or PCMCIA hard drives may also be used for high-capacity waveform and setup
storage.
CURSOR MEASUREMENTS
Relative Time: Two cursors provide time measurements with resolution of
±0.05% full-scale for unexpanded traces; up to 10% of the sampling interval for expanded
traces. The corresponding frequency value is displayed.
Relative Voltage: Two horizontal bars measure voltage differences up to
±0.2% of full-scale in single-grid mode.
Absolute Time: A cross-hair marker measures time relative to the trigger
and voltage with respect to ground.
Absolute Voltage: A reference bar measures voltage with respect to
ground.
AUTOMATIC MEASUREMENTS
The following Parametric measurements are available, together with their Average, Highest,
Lowest values and Standard Deviation:
| amplitude |
falltime |
peak to peak |
| area |
f 80-20% |
period |
| base |
f@level (abs) |
risetime |
| cycles |
f@level (%) |
r 20-80% |
| delay |
frequency |
r@level (abs) |
| delta delay |
maximum |
r@level (%) |
| delta t at level (abs) |
mean |
RMS |
| delta t at level (%) |
median |
std dev |
| delta t at level (t=0,abs) |
minimum |
top |
Pass/Fail testing allows any 5 items (parameters and/or masks) to be tested against
selectable thresholds. Waveform Limit Testing is performed using Masks which may be
defined inside the instrument. Any failure can initiate preprogrammed actions such as
Hardcopy, Save to internal memory, Save to mass storage device (card or disk), GPIB SRQ or
Pulse Out.
DC2D+ and DC2D- measure setup and hold times.
WAVEFORM PROCESSING
Up to four processing functions may be performed simultaneously. Functions available are:
Add, Subtract, Multiply, Divide, Negate, Identity and Summation Averaging.
Average: Summed averaging of up to 1,000 waveforms in the basic
instrument. Up to a million sweeps are possible with Option WP01.
Envelope*: Max, Min, or Max and Min values of up to one million sweeps.
ERES*: Low-Pass digital filter provides up to 11 bits vertical
resolution.
Sampled data is always available, even when a trace is turned off. Any of the above modes
can be invoked without destroying the data.
FFT*: Spectral Analysis with four windowing functions and FFT averaging.
*Envelope and ERES modes are provided in Advanced Math Package WP01, FFT is in Package
WP02.
AUTOSETUP
Pressing Autosetup sets timebase, trigger and sensitivity to display a wide range of
repetitive signals. (Amplitude 2 mV to 40 V; frequency above 50Hz; Duty cycle greater than
0.1%).
Autosetup Time: Approximately 2 seconds.
Vertical Find: Automatically sets sensitivity and offset.
PROBES
Model: One PP005 (X10, 10 M Ohm // 11 pF) probe supplied per channel.
The 9384 family is fully compatible with LeCroys range of FET Probes, which may be
purchased separately.
Probe calibration: Max 1 V into 1 M Ohm, 500 mV into 50 Ohm, frequency and amplitude
programmable, pulse or square wave selectable, rise and fall time 1 ns typical.
Alternatively, the Calibrator output can provide a trigger output or a PASS/FAIL test
output.
INTERFACING
Remote Control: All front-panel controls, as well as all internal
functions are possible by GPIB and RS-232-C.
RS-232-C Port (Standard): Asynchronous up to 19200 baud for computer/terminal control or
printer/plotter connection.
GPIB Port (Standard): (IEEE-488.1) Configurable as talker/listener for computer control
and fast data transfer
Centronics port: Hardcopy parallel interface are available as part of either floppy disk
or internal printer options.
Hardcopy: Screen dumps are activated by a front panel button or via remote control. TIFF
format is available for importing to Desktop Publishing programs. The following printers
and plotters can be used to make hardcopies: HPThinkJet, QuietJet, LaserJet, PaintJet, and
EPSON printers, HP7400 and 7500 series, or HPGL compatible plotters.
An optional internal high resolution graphics printer is also available, see
p. 73.
GENERAL
Auto-calibration ensures specified DC and timing accuracy.
Temperature: 5° to 40° C rated accuracy. 0° to 50° C operating.
Derate @ 1°C/1000 ft of altitude to 10,000 ft.
Humidity: <80%
Shock and Vibration: Meets MIL-STD-810C modified to LeCroy design
specifications and MIL-T-28800C.
Power: 90-250 VAC, 45-66 Hz, 350 W.
Battery Backup: Front panel settings maintained for two years.
Dimensions: (HWD) 8.5" x 14.5" x 16.25", (210mm x 370mm x
410mm).
Weight: 13 kg (28.6 lbs) net, 18.5 kg (40.7 lbs) shipping.
Warranty: Three years
Compliance: Complies with the EU - EMCDirective
|