LA302 - 100 MHz
LA303 - 200 MHz
Main
Features
-
100
& 200 MHz
-
2
+ 1 Channels
-
Sensitivity
of 2 mV/div max
-
Max
Sweep Speed of 1 ns/div
-
High-Speed
Auto Setup
-
Cursor
Measurements
-
Full
TV Trigger with Field & Line Selection
-
Save
and Recall Panel Settings
-
Frequency
Counter
-
High-Intensity
CRT
With three channels and up to 200 MHz bandwidth, the LA302 and LA303 analog oscilloscopes offer a high level of performance for relatively low cost.
Analog oscilloscopes offer unique benefits in solving specific measurement problems. The analog display provides important clues about the relative frequency content of signals mixed together or the occurrence of low rep-rate events on repetitive signals. The analog display effectively displays these "grey scaling" and "persistence" effects. The
ultra fast display update rate lets you see how the waveforms behave in real time.
The LA302 and LA303 are perfect instruments for service and educational applications. Equipped with extensive TV triggering, both scopes are excellent for video testing.
The standard multipurpose trigger, five-digit counter, and comprehensive cursors make the LA302 and LA303 truly universal oscilloscopes.
Features
Convenient
fine-adjustment mode
Pressing
FINE and turning the knob enables slow scrolling of
waveforms a function comparable to a digital
storage scope's roll mode.
Line
triggering provided for AC line measurement
AC
line-related measurement of various frequencies can
be obtained.
Third
channel input
CH3
provides three sensitivity levels for a wider range
of measurements.
Input
with probe sensor function
When
using a probe, the range display is automatically
calibrated by detecting 10:1 or 100:1 with the probe
sensor.
Specifications
|
Model LA302 |
Model
LA303 |
| Display |
CRT:
6-inch rectangular, internal graticule (8 x 10 div
with scale illumination), meshless CRT |
| Accelerating
voltage: |
Nominal 16 kV
|
| Vertical
Deflection System |
| Mode:
|
CH1, CH2, CH3, ADD, ALT, CHOP |
| Frequency
Response: |
5 mV/div - 5 V/div, DC 100 MHz; 2 mV/div, DC
50 MHz |
5 mV/div - 5 V/div, DC 200 MHz;
2 mV/div, DC 50 MHz |
| Accuracy: |
±2% |
| Bandwidth
Limiter: |
20 MHz |
| Risetime: |
3.5 ns or less |
1.75 ns or less |
| Signal
Delay Time: |
At least 30 ns |
| Input
RC Direct: |
1 MOhm ±1.5% // 20 pF ±2 pF |
| Input
RC with Probe: |
10 MOhm ±3% // 12.5 pF ±2 pF |
| Max.
Input: |
Direct, ±400 V max, with 10:1 probe, ±600
V max |
Polarity
Switching:
|
Possible only for CH2 |
| Channel
1 & 2 |
|
|
| Sensitivity:
|
2 mV/div - 5 V/div, 11 step (1-2-5); 2 mV/div - 12.5
V/div (with vernier) |
| Input
Coupling: |
AC, DC, GND (The lowest usable
frequency is 4 Hz with AC coupling.) |
| Channel
3 |
|
|
| Sensitivity:
|
50 mV/div, 100 mV/div, 500 mV/div, ±2% |
| Input
Coupling: |
AC, DC |
| Triggering |
|
|
| A
Triggering |
|
|
Sources: CH1, CH2, CH3, LINE
Coupling: AC, DC, HF-REJ, LF-REJ |
|
| B
Triggering |
|
|
Sources: CH1, CH2, CH3
Coupling: AC, DC, HF-REJ, LF-REJ |
|
| TV
Triggering: |
ODD,
EVEN, BOTH, line selection, TV system (NTSC,
PAL/SECAM [HDTV optional in LA302, standard in
LA303]) |
| Trigger
Sensitivity |
DC - 10 MHz, 0.4 div
10 - 100 MHz, 1.0 div |
DC - 10 MHz, 0.4 div
10 - 100 MHz, 1.0 div
100 - 200 MHz, 1.5 div |
X-Y
Operation
X-axis:
Sensitivity, same as CH1; accuracy, ±3%; bandwidth,
DC - 2 MHz
Y-axis:
CH1, CH2, CH3, ADD
Horizontal
Deflection System
Horizontal
Display A, ALT, B, X-Y
Mode
AUTO, NORM, SINGLE
A
sweep
Model LA302
Sweep time: Fastest range, 20 ns/div; slowest range,
500 ms/div
Model
LA303
Sweep time: Fastest range, 10 ns/div (1-2-5 step);
slowest range, 500 ms/div
B
sweep
Model LA302
Sweep time: Fastest range, 20
ns/div; slowest range, 5 ms/div; accuracy A/B, ±2%
Model
LA303
Sweep time: Fastest range, 10 ns/div (1-2-5 step);
slowest range, 5 ms/div; accuracy A/B, ±2%
Magnifier:
10 times
Accuracy:
±3%
Model
LA302
Fastest Range: 2 ns/div
Model
LA303
Fastest Range: 1 ns/div
Delay
Jitter: 1/20,000
Hold-off
Time: Variable (up to sweep length or
longer)
Auto
Setup
Input
Channel: CH1, CH2
Frequency
Response: 50 Hz 50 MHz
CH2
Signal Output
(factory
option)
Model
LA302
DC
50 MHz
Model
LA303
DC
100 MHz
EXT
Z-axis Input
(factory
option)
0.5
Vp-p
Utilities
Calibrator
Waveform:
Square-wave, 1 kHz ±0.1%, 0.6 Vp-p ±1%
CTR
Readout
Readout:
Attenuator range, ADD, UNCAL, AC/DC/GND, vertical
mode, CH2 polarity, A/B sweep range, x10 MAG, UNCAL,
horizontal display mode, hold-off, trigger coupling,
trigger source, trig slope, TV-field, TV-line,
TV-system (NTSC, PAL/SECAM [HDTV in LA303])
Cursor
Measurement
Time
difference: DeltaT
Voltage
difference: DeltaV
Division
Calculation by DeltaT: 1/DeltaT
Frequency
Counter
Display
digits: 5 digits, decimal
Accuracy:
±0.01%
Model
LA302
Frequency measurement range:
2 Hz - 100 MHz (option six digit)
Model
LA303
Frequency measurement range:
2 Hz - 200 MHz, reciprocal (option six digit)
Save/Recall
Function
Number
of Panel Setups:
Model LA302 and LA303
32 max
Model
LA302P
255 in banks of 63
Back-up:
Battery back-up (approx. 30,000 hr)
Power
Supply
Voltage
range: AC 90 V - 132 V/ 180 V - 250 V
Frequency
range: 48 Hz - 440 Hz
Power
consumption: 110 VA max
Dimensions
and Weight
Dimensions:
272W x 152H x 410L mm
Weight:
approx. 7.5 kg (w/o access)
Environmental
Conditions
Operating:
0°C +40°C, 90% RH (40°C)
Performance
Guarantee: +10°C +35°C
|